Low-profile Two-axis, Positioning with Clear Aperture

The Dover OFSā„¢ series inverted microscope XY stage combines low-profile, two-axis positioning with a clear aperture, allowing it to be used for applications requiring transmitted light, cable, or pneumatic access through the bottom.

Specifications

Inverted Microscope XY Stage ā€“ OFS


wdt_ID Base Model OFS-4040 OFS-6060 OFS-8080 OFS-1010
1 Travel (mm) 50x50 100x100 150x150 200x200
2 Travel (in) 2 x 2 4 x 4 6 x 6 8 x 8
3 Accuracy (Ī¼m TIR) Commercial Grade 45 65 75 80
4 Bi-directional Repeatability (Ā± Ī¼m) Commercial Grade 5 5 5 5
8 Accuracy (Ī¼m TIR) Precision Grade 15 20 20 20
9 Bi-directional Repeatability (Ā± Ī¼m) Precision Grade 2.5 2.5 2.5 2.5
10 Load Capacity (kg) 2 5 5 5
11 Flatness & Straightness (Ī¼m TIR) 5 10 15 20
12 Total Mass (kg) 2 3 5 8

Print Datasheet

Description

The OFSā„¢ series has orthogonality ā€œmachined-inā€ by the three-piece construction of theseĀ XY stagesĀ that employ ball ways. Both motors are attached to the center section so that they both move with the lower axis and the cabling only flexes in one direction of travel. The 1.5-inch-by-1.5-inch clear aperture allows OFSā„¢ series tables to be used in applications requiring transmitted light, cable or pneumatic access through the bottom, or two-sided access to the userā€™s part. The use of side-mounted lead screws with spring-loaded friction nuts driven by 17-frame motors achieves a stage body thickness of only 1 inch. The motors include rear-shaft mounted knobs for manual control, and locking motor and limit/encoder connectors allow convenient attachment of cabling.

Design Features

  • Clear aperture for two-sided access
  • Large 42 millimeter unobstructed opening

Options

  • Motor options available: stepper motors or servo motors
  • Feedback options available: rotary encoder, linear encoder with analog or digital (1 um or 0.1 um resolution are standard, other resolutions available upon request)
  • Download the datasheet for this linear motion XY stage for additional technical details

Applications

The DoverĀ OFSā„¢ inverted microscope XY stage has been integrated with microscopes to allow backlighting of a sample from below.

Downloads

Datasheets

CAD Downloads